Free · 13,591 studies indexed

lighttherapyresearch.com is a free index of research on photobiomodulation and light therapy, gathered from the world’s journals and trial registries.

Sections
Devices & Dosimetry

Design and testing of low intensity laser biostimulator

The prototype's specifications were found to be better or comparable to those of other existing devices. It features low component count, small size and low power consumption.

Excerpt from the paper’s abstract. Read the full paper for the details.

AI-generated summary

The authors describe the design and bench testing of a low-cost portable laser device for stimulating acupuncture points, using a 685 nm diode at low power. In a prototype test on one meridian, they report recording periodic evoked electrical potentials at accessible sites. The authors state that more studies are needed before any general conclusion about the stimulation mechanism or effective settings can be drawn.

Who or what was studied
unclear
Wavelengths
685 nm
Limits stated in the abstract
The abstract describes only a prototype test on one meridian, with no reported participant count, control group or comparison, and the authors state more studies are needed.

Written by an AI model from this paper’s abstract only and checked automatically against it, but it can still be wrong. It is not medical advice. Read the full paper. See the editorial policy.

Abstract

Background

The non-invasive nature of laser biostimulation has made lasers an attractive alternative in Medical Acupuncture at the last 25 years. However, there is still an uncertainty as to whether they work or their effect is just placebo. Although a plethora of scientific papers published about the topic showing positive clinical results, there is still a lack of objective scientific proofs about the biostimulation effect of lasers in Medical Acupuncture. The objective of this work was to design and build a low cost portable laser device for stimulation of acupuncture points, considered here as small localized biosources (SLB), without stimulating any sensory nerves via shock or heat and to find out a suitable method for objectively evaluating its stimulating effect. The design is aimed for studying SLB potentials provoked by laser stimulus, in search for objective proofs of the biostimulation effect of lasers used in Medical Acupuncture.

Methods

The proposed biostimulator features two operational modes: program mode and stimulation mode and two output polarization modes: linearly and circularly polarized laser emission. In program mode, different user-defined stimulation protocols can be created and memorized. The laser output can be either continuous or pulse modulated. Each stimulation session consists of a pre-defined number of successive continuous or square pulse modulated sequences of laser emission. The variable parameters of the laser output are: average output power, pulse width, pulse period, and continuous or pulsed sequence duration and repetition period. In stimulation mode the stimulus is automatically applied according to the pre-programmed protocol. The laser source is 30 mW AlGaInP laser diode with an emission wavelength of 685 nm, driven by a highly integrated driver. The optical system designed for beam collimation and polarization change uses single collimating lens with large numerical aperture, linear polarizer and a quarter-wave retardation plate. The proposed method for testing the device efficiency employs a biofeedback from the subject by recording the biopotentials evoked by the laser stimulus at related distant SLB sites. Therefore measuring of SLB biopotentials caused by the stimulus would indicate that a biopotential has been evoked at the irradiated site and has propagated to the measurement sites, rather than being caused by local changes of the electrical skin conductivity.

Results

A prototype device was built according to the proposed design using relatively inexpensive and commercially available components. The laser output can be pulse modulated from 0.1 to 1000 Hz with a duty factor from 10 to 90%. The average output power density can be adjusted in the range 24-480 mW/cm2, where the total irradiation is limited to 2 Joule per stimulation session. The device is controlled by an 8-bit RISC Flash microcontroller with internal RAM and EEPROM memory, which allows for a wide range of different stimulation protocols to be implemented and memorized. The integrated laser diode driver with its onboard light power control loop provides safe and consistent laser modulation. The prototype was tested on the right Tri-Heater (TH) acupuncture meridian according to the proposed method. Laser evoked potentials were recorded from most of the easily accessible SLB along the meridian under study. They appear like periodical spikes with a repetition rate from 0.05 to 10 Hz and amplitude range 0.1-1 mV.

Conclusion

The prototype's specifications were found to be better or comparable to those of other existing devices. It features low component count, small size and low power consumption. Because of the low power levels used the possibility of sensory nerve stimulation via the phenomenon of shock or heat is excluded. Thus senseless optical stimulation is achieved. The optical system presented offers simple and cost effective way for beam collimation and polarization change. The novel method proposed for testing the device efficiency allows for objectively recording of SLB potentials evoked by laser stimulus. Based on the biopotential records obtained with this method, a scientifically based conclusion can be drawn about the effectiveness of the commercially available devices for low-level laser therapy used in Medical Acupuncture. The prototype tests showed that with the biostimulator presented, SLB could be effectively stimulated at low power levels. However more studies are needed to derive a general conclusion about the SLB biostimulation mechanism of lasers and their most effective power and optical settings.

Abstract reproduced under the CC BY license. Source: PubMed, a service of the National Library of Medicine.

Light parameters mentioned

Found in the title and abstract by pattern. They may be incomplete, so check the paper before comparing studies.

Wavelength
685 nm
Power density (irradiance)
480 mW/cm2
Power
30 mW

How to read a trial: what to check before relying on this kind of research.

Terms in this paper: Laser, Low-level laser therapy, Wavelength. Each links to its definition in the glossary.

A summary of published research, not medical advice. One study rarely settles a question. See the disclaimer and how to read research.

Related studies

Devices & Dosimetry

Design of the laser acupuncture therapeutic instrument

M5P430 microcontroller system is used as the control centre of the instrument. The realization of lifting and thrusting manipulating method and twisting and twirling manipulating method are technological innovations in traditional acupuncture coming true in engineering.

Study · January 1, 2006 · Li C et al.