The authors measured the beam characteristics and optical output of one red-light myopia control device, using artificial pupils of 1 to 9 mm. They compared estimated retinal exposure to ANSI safety limits for ophthalmic instruments. They report that retinal exposure exceeded thermal limits for pupils above 1 mm and photochemical limits for pupils above 5 mm, with exposure varying with pupil size, accommodation and eye movements.
- Wavelengths
- 655 nm
- Limits stated in the abstract
- The analysis was a bench measurement of a single sample device using artificial pupils and conservative estimates, not a study in human participants.
Written by an AI model from this paper’s abstract only and checked automatically against it, but it can still be wrong. It is not medical advice. Read the full paper. See the editorial policy.
Abstract
Purpose
Recent literature has demonstrated that 'low-level red-light therapy' may be effective at slowing axial elongation in children, but reports have questioned the safety of some red-light devices. The current work explored the beam characteristics and hazard potential for a myopia control red-light device.
Methods
The optical design and exiting beam characteristics of a sample red-light myopia control device (SECONEE sky-n1201a) were quantified by measuring beam divergence and total flux passing through artificial pupils of 1-9 mm diameter placed at the corneal plane. Retinal exposure levels were compared to American National Standards Institute (ANSI) limits for ophthalmic instruments (Z80.36-2021).
Results
Two 655 nm laser diodes created exiting red-light beams (~9 mm in diameter) that were approximately collimated by an internal pair of +10.00 D eye-piece lenses resulting in exit vergences ranging from -3.25 to -7.75 D as interpupillary distance settings were adjusted from 52 to 70 mm. Radiant power (mW) passing through the artificial pupils increased from 0.005 to 0.65 mW as pupil diameters increased from 1 to 9 mm. Using ANSI Z80.36-2021, time-averaged retinal radiant exposures (J/cm2) exceeded the thermal and photochemical limits for pupil diameters >1 mm and >5 mm, respectively. The time to reach potential photochemical radiation hazard (tmax) was less than 100 s for pupil diameters >5 mm.
Conclusions
The optical design and resulting beam characteristics of this sample red-light myopia control device generate retinal exposure levels that vary with pupil diameter, accommodation and eye movements. Conservative estimates reveal retinal radiant exposures that can exceed ANSI safety limits. The 'low-level' moniker is inappropriate for a device that either exceeds or is just below the ANSI threshold for potential retinal hazard.
Abstract reproduced under the CC BY license. Source: PubMed, a service of the National Library of Medicine.
Light parameters mentioned
Found in the title and abstract by pattern. They may be incomplete, so check the paper before comparing studies.
- Wavelength
- 655 nm
- Power
- 0.65 mW
How to read a single study: what to check before relying on this kind of research.
Terms in this paper: Myopia, Laser. Each links to its definition in the glossary.
A summary of published research, not medical advice. One study rarely settles a question. See the disclaimer and how to read research.