AI-generated summary
The authors measured optical output and ANSI safety classification of four commercial red light therapy devices for myopia in a laboratory study. They report that two laser-based devices, Sky-n1201 and EyeRising, reached ANSI group 1 retinal limits within 2.8 and 1.4 seconds, respectively, for a 7-mm pupil, while the LED-based AirDoc did not reach its limit until 22 761 seconds. The authors conclude that independent safety validation is needed before widespread pediatric use.
- Limits stated in the abstract
- The study is a laboratory evaluation of four devices, not a clinical study in people, so it does not directly measure ocular harm.
Written by an AI model from this paper’s abstract only and checked automatically against it, but it can still be wrong. It is not medical advice. Read the full paper. See the editorial policy.
Abstract
The full abstract is not shown here because it is protected by the publisher’s copyright. Read it on jamanetwork.com.
Record from PubMed, a service of the National Library of Medicine.
How to read a single study: what to check before relying on this kind of research.